Background
CMR typically quantifies LV mass (LVM) via manual planimetry (MP), but this approach is time consuming and does not account for partial voxel components - myocardium admixed with blood in a single voxel. Automated segmentation (AS) can account for partial voxels, but this has not been used for LVM quantification. This study used automated CMR segmentation to test the influence of partial voxels on quantification of LVM.
Methods and Results
LVM was quantified by AS and MP in 126 consecutive patients and 10 laboratory animals undergoing CMR. AS yielded both partial voxel (ASPV) and full voxel (ASFV) measurements. Methods were independently compared to LVM quantified on echocardiography (echo) and an ex-vivo standard of LVM at necropsy. AS quantified LVM in all patients, yielding a 12-fold decrease in processing time vs. MP (0:21±0:04 vs. 4:18±1:02 min; p<0.001). ASFV mass (136±35gm) was slightly lower than MP (139±35; Δ=3±9gm, p<0.001). Both methods yielded similar proportions of patients with LV remodeling (p=0.73) and hypertrophy (p=1.00). Regarding partial voxel segmentation, ASPV yielded higher LVM (159±38gm) than MP (Δ=20±10gm) and ASFV (Δ=23±6gm, both p<0.001), corresponding to relative increases of 14% and 17%. In multivariable analysis, magnitude of difference between ASPV and ASFV correlated with larger voxel size (partial r=0.37, p<0.001) even after controlling for LV chamber volume (r=.28, p=0.002) and total LVM (r=0.19, p=0.03). Among patients, ASPV yielded better agreement with echo (Δ=20±25gm) than did ASFV (Δ=43±24gm) or MP (Δ=40±22gm, both p<0.001). Among laboratory animals, ASPV and ex-vivo results were similar (Δ=1±3gm, p=0.3), whereas ASFV (6±3gm, p<0.001) and MP (4±5gm, p=0.02) yielded small but significant differences with LVM at necropsy.
Conclusions
Automated segmentation of myocardial partial voxels yields a 14-17% increase in LVM vs. full voxel segmentation, with increased differences correlated with lower spatial resolution. Partial voxel segmentation yields improved CMR agreement with echo and necropsy-verified LVM.
CMR quantification of LV chamber volumes typically and manually defines the basal-most LV, which adds processing time and user-dependence. This study developed an LV segmentation method that is fully automated based on the spatiotemporal continuity of the LV (LV-FAST). An iteratively decreasing threshold region growing approach was used first from the midventricle to the apex, until the LV area and shape discontinued, and then from midventricle to the base, until less than 50% of the myocardium circumference was observable. Region growth was constrained by LV spatiotemporal continuity to improve robustness of apical and basal segmentations. The LV-FAST method was compared with manual tracing on cardiac cine MRI data of 45 consecutive patients. Of the 45 patients, LV-FAST and manual selection identified the same apical slices at both ED and ES and the same basal slices at both ED and ES in 38, 38, 38, and 41 cases, respectively, and their measurements agreed within −1.6 ± 8.7 mL, −1.4 ± 7.8 mL, and 1.0 ± 5.8% for EDV, ESV, and EF, respectively. LV-FAST allowed LV volume-time course quantitatively measured within 3 seconds on a standard desktop computer, which is fast and accurate for processing the cine volumetric cardiac MRI data, and enables LV filling course quantification over the cardiac cycle.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.