A method for the determination of nickel(II) by stripping voltammetry is described. The method is based on the adsorptive accumulation of nickel(II) calconcarboxylic acid complex onto a hanging mercury drop electrode (HMDE), followed by the reduction of the adsorbed complex using differential pulse voltammetry. The optimum operating conditions and parameters were found to be 0.05 M NH 3 /NH 4 Cl buffer (pH = 9.5) as the supporting electrolyte, a ligand concentration of 1 × 10 -6 M, accumulation potential of -0.5 V (vs. Ag/AgCl) and accumulation time of 60 s. At the optimized conditions, the peak current is proportional to the concentration of nickel in the range of 1.7 × 10 -9 to 4.7 × 10 -7 M (0.1-28 ng ml -) with a detection limit of 0.05 ng ml -1 . The relative standard deviation (n = 10) at nickel concentrations of 2, 10 and 15 ng ml -1 varies in the range 0.76 to 2.1%. Possible interferences by metal ions, which are of great significance in real matrices, have been studied. The method was successfully applied to the determination of nickel content in a chocolate sample.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.