Calibration of axial quadrant photodetector (QPD) signal to the trapped bead position in an optical tweezer is important to measure the quantitative mechanical parameter in axial (laser propagation) direction. An alternative calibration based on the Stuck Bead Method (SBM) was proposed in this study. 3 µm polystyrene beads were stuck at the surface of glass coverslip and moved axially around the laser focus. QPD was used to obtain the position dependent intensity profile at three different laser powers (19.8 mW, 34.1 mW, 48.5 mW). The QPD signal-to-distance calibration value was consistent at 26 mV/µm for the used bead at the three laser powers. It was found that the calibration values are independent of laser powers and limited by the resolution of distance adjustment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.