We have developed a Threshold System immunoligand assay for the quantitation of residual, process-specific, Escherichia coli host cell contaminant proteins (HCP) in somavubove (a normal sequence recombinant bovine somatotropin). The assay has a dynamic range of 2 to 160 ng/mL, with a limit of quantitation of 2 ng/mL. Daily analytical precision (CV) for six replicates of the designated somavubove laboratory standard is 3.0%. Cumulative analytical precision for multiple assay runs of this standard (a total of 69 laboratory standard replicates) is 8.4%. A conservative alert limit of 100 ng/mL has been assigned in order to assure analytical precision for each assay sample under conditions of stoichiometric antibody excess. Although the assay was designed for use as a profile-release specification assay, it has also been used to validate removal of HCP by the proprietary somavubove purification process. This use is consistent with regulatory guidelines related to "well characterized" recombinant biopharmaceutical proteins.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.