The CardioVention system with its concept of limited priming volume and exposed foreign surface area, improves gas exchange probably because of the absence of detectable hemodilution, and appears to limit the decrease in the thrombocyte count which may be ascribed to the reduced surface. Despite the volume and surface constraints, no hemolysis could be detected throughout the 6 h full-flow perfusion period.
Although gravity drainage has been the standard technique for cardiopulmonary bypass (CPB), the development of min imally invasive techniques for cardiac surgery has renewed interest in using vacuum assisted venous drainage (VAVD) Dideco (Mirandola, Italy) has modified the D903 Avant oxygenator to apply a vacuum to its venous reservoir. The impact of VAVD on blood damage with this device is analyzed. Six calves (mean body weight, 71.3 +/- 4.1 kg) were con nected to CPB by jugular venous and carotid arterial cannu lation, with a flow rate of 4-4.51 L/min for 6 h. They were assigned to gravity drainage (standard D903 Avant oxygen ator, n = 3) or VAVD (modified D903 Avant oxygenator, n = 3). The animals were allowed to survive for 7 days. A standard battery of blood samples was taken before bypass, throughout bypass, and 24 h, 48 h, and 7 days after bypass. Analysis of variance was used for repeated measurements. Thrombocyte and white blood cell counts, corrected by hematocrit and normalized by prebypass values, were not significantly different between groups throughout all study periods. The same holds true for hemolytic parameters (lactate dehydrogenase [LDH] and plasma hemoglobin). Both peaked at 24 hr in the standard and VAVD groups: LDH, 2,845 +/- 974 IU/L vs. 2,537 +/- 476 IU/L (p = 0.65), respectively; and plasma hemoglobin, 115 +/- 31 mg/L vs. 89 +/- 455 mg/L (p = 0.45), respectively. In this experimental setup with prolonged perfusion time, VAVD does not increase trauma to blood cells in comparison with standard gravity drainage.
When compared with uncoated oxygenator, PMEA coated oxygenator exhibited increased thrombus resistance with lower inlet pressure and lower thrombocyte consumption. In both groups, trauma to red cells was minimal, emphasizing the efficient design of this type of oxygenator.
In this experimental set-up, the Smart suction system avoids damage to red cells and affects platelet count less than the Cell Saver system which induces important blood cell destruction, as any suction device mixing air and blood, as well as severe hypoproteinemia with its metabolic, clotting and hemodynamic consequences.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.