The reproducibility of the Lenstar was excellent. Small but significant differences exist between the Lenstar, Visante OCT and the IOLMaster. Therefore, measurements of the Lenstar, AS-OCT and IOLMaster are not interchangeable. Despite the significant difference between AL measurements, there is no clinically significant difference in the IOL power calculation results.
BackgroundThe repeatability and interchangeability of imaging devices measuring central corneal thickness (CCT) and anterior chamber depth (ACD) are important in the assessment of patients considering refractive surgery. The purpose of this study was to investigate the agreement of CCT and ACD measurements using three imaging technologies in healthy eyes and in eyes after phakic intraocular lens implantation (pIOL).MethodsIn this comparative study, CCT and ACD were measured using anterior segment optical coherence tomography (AS-OCT), Orbscan II, and Pentacam in 33 healthy volunteers (66 eyes) and 22 patients (42 eyes) after pIOL implantation. Intraobserver repeatability was evaluated for all three devices in the healthy volunteer group.ResultsPairwise comparison of CCT measurements showed significant differences between all devices (P < 0.001), except for the AS-OCT and Orbscan II in the healthy volunteer group (P = 0.422) and the Orbscan II and Pentacam in the pIOL group (P = 0.214). ACD measurements demonstrated significant differences between all pairwise comparisons in both groups (P ≤ 0.001). Intraobserver reliability was high for CCT and ACD measurements in the healthy volunteer group, with coefficients of variation ranging from 0.6% to 1.2% and 0.4% to 0.5% respectively.ConclusionsCCT and ACD measurements using AS-OCT, Orbscan II, and Pentacam demonstrated high intraobserver reliability. However, these devices should not be used interchangeably for measurements of CCT and ACD in healthy subject and patients after pIOL implantation.
A shorter distance between the edge of the pIOL and the endothelium was significantly associated with higher ECD loss. For safety reasons, the postoperative examination should include long-term evaluation of the anterior chamber morphometrics in addition to ECD counts.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.