Irregular breathing in lung cancer patients is a common contraindication to 4D computerized tomography (4DCT), which may then limit radiotherapy treatment options. For irregular breathers, we investigated whether 3DCT or 4DCT (1) better represents tumor motion, (2) better represents average tumor densities, and (3) better allows for volumetric modulated arc threarpy (VMAT) plans delivered with acceptable dosimetric accuracy. Methods: Ten clinical breathing traces were identified with irregularities in phase and amplitude, and fed to a programmable moving platform incorporating an anthropomorphic lung tumor phantom. 3DCT and 4DCT data resorted by phase (4DCT-P) and amplitude (4DCT-A) were acquired for each trace. Tumors were delineated by Hounsfield unit (HU) thresholding and apparent motion range assessed. HU profiles were extracted from each image and agreement with calculated expected profiles quantified using area-under-curve (AUC) scoring. Clinically representative VMAT plans were created for each image, delivered to the irregularly moving phantom, and measured with a small-volume ion chamber at the tumor center. Results: Median difference from expected tumor motion range for 3DCT, 4DCT-P, and 4DCT-A was 2.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.