Dedicated nitric oxide equipped ventilators are now available commercially but are not yet common in clinical practice. With other ventilators, there is no standardized procedure for the administration or monitoring of nitric oxide. We describe the use of nitric oxide in conjunction with a simple time-cycled, pressure regulated, flow generating ventilator attached to a model infant-sized lung. The measured nitric oxide concentrations were always less than calculated. Infusion site, minute ventilation and sampling port all affected nitric oxide concentration (P < 0.05). Increasing minute ventilation lowered measured nitric oxide concentration exponentially. Mixing of gases improved when nitric oxide was infused closer to the ventilator. Acid contamination was found in water samples from humidifier, water trap and ventilator gas outlet. Acidification was reduced, without change in measured nitric oxide delivery, when infused prehumidifier. We recommend, when used as therapy, nitric oxide levels in inspired gases should always be measured.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.