The use of x-ray computed tomography for dimensional measurements associated with engineering applications has flourished in recent years. However, error sources associated with the technology are not well understood. In this paper, a novel two-sphere reference sample has been developed and used to investigate the stability of the imaging system that consists of an x-ray tube and a detector. In contrast with other research work reported, this work considered relative positional variation along the x-, y- and z-axes. This sample is a significant improvement over the one sphere sample proposed previously, which can only be used to observe the stability of the imaging system along x- and y-axes. Temperature variations of different parts of the system have been monitored and the relationship between temperature variations and x-ray image stability has been studied. Other effects that may also influence the stability of the imaging system have been discussed. The proposed reference sample and testing method are transferable to other types of x-ray computed tomography systems, for example, systems with transmission targets and systems with sub-micrometre focal spots.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.