Objective
The introduction of ablation technology has simplified surgical intervention for atrial fibrillation. However, most ablation devices cannot create focal transmural lesions on the beating heart and have difficulty ablating specific regions of the atria, such as the atrioventricular isthmus, coronary sinus, and ganglionated plexus. The purpose of this study was to examine the efficacy of a pen-type bipolar radiofrequency ablation device on both arrested and beating hearts.
Methods
Endocardial and epicardial atrial tissues in the free wall, left atrial roof, atrioventricular annuli, and coronary sinus were ablated for varying time intervals (2.5–15 seconds) in porcine cardioplegically arrested (n = 6) and beating (n = 9) hearts. The hearts were stained with 1%2,3,5-triphenyl-tetrazolium chloride solution and sectioned to determine lesion depth and width. In 5 animals epicardial fat pads containing ganglionated plexus were stimulated and ablated.
Results
Lesion depth increased with ablation time similarly in both arrested and beating hearts. Transmurality was fully achieved in the thin atrial tissue (<4 mm) at 10 seconds in the beating and arrested hearts. The device had a maximal penetration depth of 6.1 mm. Epicardial ablation of the coronary sinus showed complete penetration through the left posterior atrium only in the arrested heart. Seven of 17 fat pads demonstrated a vagal response. All vagal responses were eliminated after ablation.
Conclusion
The bipolar pen effectively ablated atrial tissue in both arrested and beating hearts. This device might allow the surgeon to ablate tissue in regions not accessible to other devices during atrial fibrillation surgery.
Cardiogenic shock (CS) is a life-threatening condition associated with significant morbidity and mortality. The Impella (Abiomed Inc.) is an axial flow pump on a pigtail catheter that is placed across the aortic valve to unload the left ventricle by delivering non-pulsatile blood flow to the ascending aorta. It is used for high-risk percutaneous coronary intervention and CS. Areas covered: Percutaneous mechanical support devices are placed in a minimally invasive manner and provide life-saving assistance. We review Impella and other percutaneous devices such as intra-aortic balloon pump, TandemHeart, and extracorporeal membrane oxygenation (ECMO) and the evidence supporting their use in the setting of CS. Expert commentary: Impella has been proven to be safe and may be superior to other mechanical support devices in CS.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.