Abstract. Selenium (Se) is an essential trace element that is often deficient in the natural diets of domestic animal species. The measurement of Se in whole blood or liver is the most accurate way to assess Se status for diagnostic purposes. This study was conducted to compare hydride generation atomic absorption spectroscopy (HG-AAS) with inductively coupled plasma-mass spectrometry (ICP-MS) for the detection and quantification of Se in liver samples. Sample digestion was accomplished with magnesium nitrate and nitric acid for HG-AAS and ICP-MS, respectively. The ICP-MS detection was optimized for 82 Se with yttrium used as the internal standard and resulted in a method detection limit of 0.12 g/g. Selenium was quantified by both methods in 310 samples from a variety of species that were submitted to the Toxicology Laboratory at New Bolton Center (Kennett Square, PA) for routine diagnostic testing. Paired measurements for each sample were evaluated by a mean difference plot method. Limits of agreement were used to describe the maximum differences likely to occur between the 2 methods. Results suggest that under the specified conditions ICP-MS can be reliably used in place of AAS for quantitation of tissue Se at or below 2 g/g to differentiate between adequate and deficient liver Se concentrations.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.