Dansyl-DL-amino acids were separated by tube radial distribution chromatography (TRDC) comprising a commercially available HPLC system, where open-tubular capillary tubes manufactured for capillary GC were used as the separation column. Initially, the separation performance of GC capillary tubes in TRDC was assessed by separating a model analyte mix of 1-naphthol and 2,6-naphthalenedisulfonic acid by using water/acetonitrile/ethyl acetate solutions (3:8:4 volume ratio, organic solvent-rich and 4:3:1 volume ratio, water-rich) as the eluent and high-polarity and non-polarity capillary tubes. Next, enantiomer separation of dansyl-DL-methionine and dansyl-DL-valine was examined using a water/acetonitrile/ethyl acetate solution (8:2:1 volume ratio, water-rich) containing 1.0 mM -cyclodextrin and non-polarity capillary tubing. The D-and L-enantiomers were separated and detected in this order through interaction between amino acids and -cyclodextrin by tube radial distribution flow in the TRDC system.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.