Three pediatric patients presented with histologically confirmed osteolysis after limb lengthening with a magnetic, telescoping, stainless-steel device. The first patient's findings were discovered radiographically before routine removal of the device. In all cases, intraoperative histologic specimens taken from around the modular junction demonstrated particle-laden macrophages with suspicion for metal debris. Silicone debris was also identified.
We found definitive osteolysis secondary to metal at the modular junction of three stainless-steel lengthening implants. This process is not well-understood in the setting of limb lengthening and should be examined further.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.