In this study, tetra-substituted copper(II) phthalocyanine-graphene modified pencil graphite electrode (CuPc-Gr/PGE) was developed for the amperometric determination of hydrogen peroxide (H2O2) for the first time in the literature. The developed non-enzymatic electrode platform was electrochemically characterized by cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) methods. Scanning electron microscopy (SEM) and X-ray photoelectron spectroscopy (XPS) methods were also used for surface characterization of the modified electrode. The response to H2O2 at the working potential of −0.4 V is linear in the 0.1 μM to 100 μM concentration range in PBS pH 7.0. The limit of detection (LOD) and limit of quantitation (LOQ) are 0.045 and 0.15 μM, respectively. Good selectivity was observed within physiological interference species such as ascorbic acid, dopamine, uric acid, and glucose. Additionally, amount of H2O2 was determined using the novel produced non-enzymatic sensor with satisfactory results in beverage samples. These consequences imply that the CuPc-Gr/PGE displays effective electrocatalytic activity for the reduction of H2O2, indicating it has capability use for the detection of H2O2.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.