SummaryTo avoid potentially fatal wrong‐route neuraxial drug errors, international standard ISO 80369‐6 specifying a non‐Luer neuraxial connector design was published in 2016. We describe usability studies used in development of the design. Thirty‐eight doctors and 17 nurses performed simulated procedures on manikins, using devices fitted with Luer connectors or draft ISO 80369‐6 ‘non‐Luer’ connectors. The procedures included spinal anaesthesia; intrathecal chemotherapy; lumbar puncture, cerebrospinal fluid collection and pressure measurement; epidural catheter placement with bolus injection and critical care use. Participants attempted cross connection between neuraxial connectors and a range of other medical device connectors, including those from the ISO 80369 small‐bore connector series. Video recording analysis was used for all assessments. Participants subjectively assessed performance of the draft non‐Luer connector, including suitability for routine clinical use. Participants performed 198 procedures. The connector achieved easy, leak‐free connections. The willingness of participants to use the non‐Luer connectors were: spinal anaesthesia 100%; intrathecal chemotherapy 88%; lumbar puncture, cerebrospinal fluid collection and pressure measurement 93%; epidural catheter placement with bolus injection 78%; critical care use 100%. Concerns raised were generally device related, rather than connector related. Most cross‐connection attempts failed, even using above clinical forces and, when successful, were judged of low clinical risk potential; the exception was a malaligned connection between the non‐Luer slip and female Luer connectors. This led to revision of the dimensional tolerances of the non‐Luer connector to reduce this risk, before publication of the final specification in 2016. We conclude that the ISO 80369‐6 neuraxial non‐Luer connector is suitable for clinical use.
(Anaesthesia. 2019;74(5):619–629) The medical literature has reported incidents of “wrong route” administration of medicines, with much concern around errors involving neuraxial administration of intravenous agents. The risk of these errors increases when devices are fitted with the “universal” Luer connector. As a result, there have been efforts to introduce non-Luer neuraxial connectors into clinical practice. This article described a study conducted to assess the clinical usability and cross-connectability of non-Luer neuraxial connectors designed for anesthesia procedures. The study was conducted as part of the development phase of the International Organization for Standardization (ISO) 80369 standard, which included specifications for a series of 7 small-bore connectors of varying sizes designed to reduce the risk of accidental misconnections.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.