A mass sensor innovative concept is presented here, based on a hollow plate Micro Electro Mechanical System (MEMS) resonator. This approach consists in running a solution through an embedded microchannel, while the plate resonator is actuated according to a Lamé-mode by electrostatic coupling in dry environment. The experimental results have shown a clear relationship between the measured shift of the resonance frequency and the sample solution density. Additionally, depending on the channel design and the solution properties, the quality factor (Q-factor) was noticed maintaining its level and even substantial improvement in particular cases. Resonators demonstrate resonance frequencies close to 78 MHz and Q-factor of a few thousands for liquid phase detection operating at ambient temperature and atmospheric pressure. Frequency fluctuations study revealed a 13 Hz instability level, equivalent to 1.5 fg in mass. Using a fully electronic readout configuration, a mass responsivity of ca. 850 fg kHz(-1) was monitored.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.