In the metrology area, typical stabilized lasers are used as length primary standard. KIM-1, i.e. the iodine stabilized He-Ne laser in National Measurement Standards Laboratory-National Standardization Agency of Indonesia (SNSU-BSN) has been traceable to SI through CCL-K11 inter-laboratory comparison. The result was suitable for KIM-1 to be used as the length primary standard in SNSU-BSN. Beat frequency measurement has been applied in the optical frequency and wavelength calibration system for stabilized He-Ne laser 633 nm using KIM-1 as a reference standard. In the calibration replica, a dual-frequency mode (Agilent 5519B) took a role as DUT laser, which emits a pair of beams with a central wavelength of 632.991 354 nm in the vacuum with frequency difference 3.4 to 4.0 MHz and ±0.02 ppm stability for a typical lifetime. As a measurement result, the beat frequencies of the 1st and 2nd polarized beam of Agilent 5519B against KIM-1 are (121.33±0.06) MHz and (118.59±0.06) MHz.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.