A reversed‐phase high‐performance liquid chromatography (RP‐HPLC) method was developed and validated for the identification and quantification of moxifloxacin hydrochloride–related substances in finished dosage forms. Chromatographic separation was achieved on an Agilent C18 column (150 × 4.6 mm, 5 μm) with the mobile phase (0.01 M potassium dihydrogen orthophosphate as buffer and methanol in the ratio of 70:30) eluted in isocratic mode. The HPLC flow rate was 1.0 mL min−1, and peaks were monitored at 230 nm using a photodiode array (PDA) detector. The column temperature was kept constant at 30°C, and the injection volume was 10 μL. The run time of the method was 16 min. The method was validated as per the International Conference on Harmonization (ICH) guidelines. Linearity was recorded at various concentrations ranging from 0.25 to 1.5 μg mL−1 for all the moxifloxacin impurities. Linearity, regression value, recovery, %relative standard deviation (RSD) of method precision values were found within the acceptance limits. The method for related substances (RS) in moxifloxacin was found to be specific, linear, accurate, precise, rugged, and robust. The validated method was suitable for the quantification of the RSs in moxifloxacin drug products. The method was applied in quality control lab for the analysis of moxifloxacin impurities in stability analysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.