An accurate detection of the radiation field is crucial to 3D conformal radiotherapy (3D‐CRT). Automated techniques to detect the field edges on double exposure portal images have previously focused on thresholding techniques. In this paper, we present a new approach based on a curve propagation technique (the Fast Marching method) which proves to be more effective at detecting the radiation field than its thresholding counterpart. The comparison of both techniques in terms of computational speed and effectiveness of the detection is presented using complex images with non‐homogeneous intensity levels inside the radiation field, and gradual variations in intensity level at the field boundaries. Results show that our Fast Marching method is easier to automate, and converges faster to the boundaries of the segmented radiation field. The computation time of the Fast Marching technique is five times faster in typical portal images.PACS numbers: 87.53.Oq, 87.57.Nk, 87.57.‐s.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.