Two multiresidue methods based on different extraction procedures have been developed and compared for the liquid chromatography electrospray ionization tandem mass spectrometry analysis of 17 mycotoxins including ochratoxin A, aflatoxins (B(1), B(2), G(1), and G(2)), zearalenone, fumonisins (B(1) and B(2)), T-2 toxin, HT-2 toxin, nivalenol, deoxynivalenol, 3- and 15-acetyldeoxynivalenol, fusarenon-X, diacetoxyscirpenol, and neosolaniol in cereal-based commodities. The extraction procedures considered were a QuEChERS-like method and one using accelerated solvent extraction (ASE). Both extraction procedures gave similar performances in terms of linearity (r(2) > 0.98) and precision (both RSD(r) and RSD(iR) < 20%). Trueness was evaluated through participation in four proficiency tests and by the analysis of two certified reference materials and one quality control material. Satisfactory Z scores (|Z| < 2) and trueness values (73-130%) were obtained by the proposed procedures. Limits of quantification were similar by both methods and were within the 1.0-2.0 microg/kg range for aflatoxins, 0.5 microg/kg for ochratoxin A, and the 5-100 microg/kg range for all other mycotoxins tested. The QuEChERS-like method was found to be easier to handle and allowed a higher sample throughput as compared to the ASE method.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.