Approximately 17 years after a primary metal-on-metal total hip arthroplasty, a 59-year-old female developed pain, swelling, and weakness in her right hip accompanied by laboratory findings and imaging suggestive of an adverse local tissue reaction. Acetabular revision was performed to upsize the femoral head and improve hip stability. Upon impaction of the new, non-option ceramic femoral head onto the unsleeved retained stem, the head split into two pieces without fragmentation. The surgery was completed using a cobalt-chromium head, which was impacted without issue onto the stem's taper. Although BIOLOX delta femoral heads do not require titanium sleeves, we believe that careful consideration should be given to their use in revision total hip arthroplasty with ceramic heads, regardless of the extent of trunnion damage noted intraoperatively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.