Therapeutic monoclonal antibodies (MAbs) are an important class of drugs used to treat diseases ranging from autoimmune disorders to B cell lymphomas to other rare conditions thought to be untreatable in the past. Many advances have been made in the characterization of immunoglobulins as a result of pharmaceutical companies investing in technologies that allow them to better understand MAbs during the development phase. Mass spectrometry is one of the new advancements utilized extensively by pharma to analyze MAbs and is now beginning to be applied in the clinical laboratory setting. The rise in the use of therapeutic MAbs has opened up new challenges for the development of assays for monitoring this class of drugs. MAbs are larger and more complex than typical small-molecule therapeutic drugs routinely analyzed by mass spectrometry. In addition, they must be quantified in samples that contain endogenous immunoglobulins with nearly identical structures. In contrast to an enzyme-linked immunosorbent assay (ELISA) for quantifying MAbs, mass spectrometry-based assays do not rely on MAb-specific reagents such as recombinant antigens and/or anti-idiotypic antibodies, and time for development is usually shorter. Furthermore, using molecular mass as a measurement tool provides increased specificity since it is a first-order principle unique to each MAb. This enables rapid quantification of MAbs and multiplexing. This review describes how mass spectrometry can become an important tool for clinical chemists and especially immunologists, who are starting to develop assays for MAbs in the clinical laboratory and are considering mass spectrometry as a versatile platform for the task.
BACKGROUND
The accurate and precise measurement of urinary albumin is critical, since even minor increases are diagnostically sensitive indicators of renal disease, cardiovascular events, and risk for death. To gain insights into potential measurement biases, we systematically compared urine albumin measurements performed by LC-MS, a clinically available immunoturbidimetric assay, and size-exclusion HPLC.
METHODS
We obtained unused clinical urine samples from 150 patients who were stratified by degrees of albuminuria (<20 mg/L, 20–250 mg/L, >250 mg/L) as determined by the immunoturbidimetric assay used in our clinical laboratory (Roche Hitachi 912). Urine albumin was then remeasured via LC-MS and HPLC (Accumin™) assays.
RESULTS
The immunoturbidimetric assay, calibrated using manufacturer-supplied serum-derived calibrators (Diasorin), underestimated albumin compared with LC-MS. After calibration with purified HSA, this immunoturbidimetric assay correlated well with LC-MS. HPLC overestimated albumin compared with both LC-MS and immunoturbidimetry. The current LC-MS and HPLC assays both performed poorly at concentrations <20 mg/L.
CONCLUSIONS
Efforts are needed to establish gold-standard traceable calibrators for clinical assays. LC-MS is a specific method to quantify albumin in native urine when concentrations exceed 20 mg/L, and therefore could be employed for standardization among assays.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.