A new sample stage for characterizing radioactive materials by X-ray powder diffraction was developed at the ATALANTE facility (CEA Marcoule, France) using a conventional (non-nuclearized) Bruker D8 goniometer mounted in Bragg–Brentano geometry. The setup consists of a removable, fully hermetic sample stage, with a 200 µm-thick beryllium window, that can be plugged onto a glove-box, allowing the sample to be introduced in an hermetic medium that also encapsulates the glove-box atmosphere throughout the analysis process. The whole setup is thus hermetically unplugged from the glove-box and positioned on the centre of the goniometer. No preliminary decontamination and/or decontainment of the sample is necessary. The device was developed to avoid an expensive and time-consuming nuclearization of the diffractometer while also keeping it easily accessible for maintenance. Ultimately, keeping the diffractometer out of a glove-box also limits the volume of the final nuclear wastes, and thus the removable sample stage is the only `active' part. X-ray diffraction results of two NIST standards LaB6 and α-Al2O3 as well as five actinide dioxides ThO2, UO2, NpO2, PuO2 and AmO2 are presented to show the efficiency of the setup.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.