Experimental valveless muscular blood pumps (biomechanical hearts) in goats can pump more than 1 L.min(-1), but due to a high pendulum volume, no significant flow contribution to the circulation is gained. Thus valved and valveless biomechanical hearts were compared for efficacy. Heart failure was induced in 5 adult Bore goats by repeated intracoronary embolization. A valved and balloon-equipped pumping chamber was integrated into the descending aorta, simulating standard biomechanical circulatory support. The valveless biomechanical heart supported a failing heart with a baseline cardiac output of 2,670 +/- 710 mL.min(-1) by contributing additional flow of 113 +/- 37 mL.min(-1). The biomechanical heart model incorporating an outlet valve offered an additional 304 +/- 126 mL.min(-1), and the use of 2 valves significantly enhanced pulmonary blood flow by 1,235 +/- 526 mL.min(-1). The use of 2 valves in biomechanical hearts seems to be essential to achieve adequate circulatory support. Double-valved biomechanical hearts driven by an appropriate skeletal muscle ventricle may contribute to the therapy of heart failure.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.