Objective:
Ankle foot orthosis (AFO) commonly prescribed to manage foot-drop following stroke restricts ankle mobility. Commercially available functional electrical stimulation (FES) is an expensive alternative to achieve desired dorsiflexion during swing phase of the gait cycle. An in-house cost-effective innovative solution was designed and developed to address this problem.The aim of the study was to compare spatiotemporal gait characteristics of patients with foot-drop following stroke using commercially available FES against in-house developed versatile single sensor-based FES.
Material and Methods:
Ten patients with cerebrovascular accident of at least 3 months duration and ambulant with/without AFO were recruited prospectively. They were trained with Device-1 (Commercial Device) and Device-2 (In-house developed, Re-Lift) for 7 h over 3 consecutive days with each device. Outcome measures included timed-up-and-go-test (TUG), six-minute-walk-test (6MWT), ten-meter-walk-test (10MWT), physiological cost index (PCI), instrumented gait analysis derived spatiotemporal parameters, and patient satisfaction feedback questionnaire. We calculated intraclass correlation between devices and median interquartile range. Statistical analysis included Wilcoxon-signed-rank-test and F-test (P < 0.05 was considered statistically significant). Bland Altman and scatter plots were plotted for both devices.
Results:
Intraclass correlation coefficient for 6MWT (0.96), 10MWT (0.97), TUG test (0.99), and PCI (0.88) reflected high agreement between the two devices. Scatter plot and Bland Altman plots for the outcome parameters showed good correlation between two FES devices. Patient satisfaction scores were equal for both Device-1 and Device-2. There was statistically significant change in swing phase ankle dorsiflexion.
Conclusions:
The study demonstrated good correlation between commercial FES and Re-Lift suggestive of the utility of low-cost FES device in clinical setting.
Introduction
The knee joint of a conventional knee-ankle-foot orthosis (KAFO) is locked during walking. This study aims to evaluate the effects of stance phase knee flexion on the kinematics, spatiotemporal parameters, and energy consumption of KAFO users.
Materials and Methods
A retrofitting drop lock that allows 12° of stance flexion in a single-axis knee joint was developed (called the semiflexion knee joint). Four subjects with quadriceps weakness volunteered to participate in the study. Gait analysis was conducted for the locked knee and the semiflexion knee orthosis using a calibrated eight-camera three-dimensional motion capture system. Preceding each trial, participants were given 5 days of gait training with the corresponding KAFO. Subjective feedback about the performance of the new KAFO was collected.
Results
Walking with the semiflexion KAFO did not significantly improve the speed (p = 0.161), cadence (p = 0.232), and stride length (p = 0.95) compared with walking with a locked KAFO. Although not significant (p = 0.132), trends of Physiological Cost Index (PCI) seemed to reduce in all subjects while walking with the semiflexion knee.
Conclusions
Using the KAFO with stance flexion did not significantly improve any of the locked knee gait parameters. However, there was a trend of reduction in the PCI score while walking with the semiflexion KAFO, and it also improved the ease of doing some activities of daily living.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.