(1) Background: A sustainable casting system that combines the use of a polystyrene bag, a prosthetic liner and a vacuum system was developed to reduce fabrication time while maintaining comfort for the trans-tibial prosthesis user. (2) Methods: Eight prosthetists (28.7 ± 8.25 years old) fit ten trans-tibial prosthesis wearers (46 ± 12.4 years old) with two types of total surface bearing (TSB) prostheses; a polystyrene bead (PS) prosthesis and a plaster of paris (POP) prosthesis. Duration of casting and combined mean peak pressure was measured at six locations on the residual limb using Force Sensing Resistors (FSR). A pressure uniformity score (%) was determined. Socket Comfort Scale (SCS) was also measured. (3) Results: Duration of casting for the POP method was 64.8 ± 9.53 min and 7.8 ± 2 min for the PS method, (p = 0.006). Pressure uniformity in the POP prosthesis was 79.3 ± 6.54 and 81.7 ± 5.83 in the PS prosthesis (p = 0.027). SCS in both prosthesis types were equivalent. (4) Conclusion: A rapid fit PS prosthesis was developed, with significantly shorter duration than the traditional POP method. Socket pressure uniformity was confirmed and improved in the PS method. Socket comfort was equal between the two prothesis types.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.