OBJECTIVES: BiliSpec is a low-cost spectrophotometric reader and disposable paper-based strip to quantify total serum bilirubin from several blood drops. This study was a prospective evaluation of BiliSpec in 2 neonatal wards in Malawi compared with a reference standard bilirubinometer over a large range of bilirubin and hematocrit levels. METHODS: The accuracy of BiliSpec and a transcutaneous bilirubinometer were compared with the reference standard of spectrophotometry for 475 blood samples collected from 375 subjects across a range of total serum bilirubin concentrations from 0.0 to 33.7 mg/dL. The development of error grids to assess the clinical effects of measurement differences is reported. RESULTS: BiliSpec was found to have a mean bias of −0.48 mg/dL and 95% limits of agreement of −5.09 mg/dL to +4.12 mg/dL. Results show 90.7% of BiliSpec measurements would have resulted in the same clinical decision as the reference standard, whereas 55.0% of transcutaneous bilirubin measurements would have resulted in the same clinical decision as the reference standard. CONCLUSIONS: This evaluation supports use of BiliSpec to provide accurate, low-cost, point-of-care bilirubin measurements in low-resource hospitals. Future work is needed to evaluate BiliSpec among a larger number of users.
Background Neonatal jaundice, defined as an excess of bilirubin in blood at birth or shortly thereafter, has a disproportionately high impact in low-resource settings. Three quarters of deaths from jaundice occur in sub-Saharan Africa and south Asia. Jaundice is easily treated with blue light phototherapy; however, accurate, low-cost bilirubin monitoring is essential to guide effective phototherapy treatment in low-resource settings. To address this need, we developed BiliDx, a low-cost, portable reader that quantifies bilirubin concentration in a few drops of blood applied to a lateral flow card. BiliDx measures absorbance at three wavelengths: blue (bilirubin), amber (haemoglobin interference), and red (background) and can perform measurements in less than 5 mins at the point of care, without the need for a costly centrifuge or extensive training. Here, we compare the accuracy of BiliDx against a standard laboratory bilirubinometer and a transcutaneous bilirubinometer. Methods This study took place in two central hospitals in Malawi. We included samples from neonates being treated in neonatal wards and deemed to be at risk for jaundice or who were undergoing phototherapy. Guardians provided written informed consent. We measured bilirubin using three devices: BiliDx, a bilirubinometer (Reichert UNISTAT), and a transcutaneous bilirubinometer (Drager JM-105), with a total of 475 measurements per device. Training (149 measurements) and validation (326 measurements) sets were created by ordering the first measurement per patient from lowest to highest UNISTAT bilirubin; every third measurement was selected for the training set. The training set was used to fit a model that correlated bilirubin concentration to absorbances at three wavelengths. The validation set was used to make an unbiased evaluation of the accuracy of the model. Findings We enrolled 456 neonates between August 7 and December 12, 2019; 247 of 326 (75%) of BiliDx measurements in the validation set were within 2 mg/dL of UNISTAT measurements, compared with 44 of 281 (6%) of quantitative transcutaneous measurements. No adverse events occurred. Interpretation BiliDx can provide accurate bilirubin monitoring. BiliDx is more accurate than transcutaneous bilirubinometry, a technology often endorsed for its non-invasive, portable nature. Limitations include that only ten nurses used BiliDx and only central hospitals were included. Testing of the BiliDx device is ongoing; we aim to collect enough measurements >20 mg/dL to assess the accuracy of BiliDx over the intended bilirubin concentration range of 0-35 mg/dL. Future studies will evaluate usability among a larger group of nurses and hospital settings. Funding USAID Saving Lives at Birth Award No. 7200AA18FA0016.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
customersupport@researchsolutions.com
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.
Copyright © 2025 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.