AbstractPurpose:Coring of a medication vial’s rubber stopper has been reported as a major cause of visible particle presence in injectable preparations. In this study, we investigated and quantified visible particle formation caused by coring associated with four potential causal factors.Methods:The factors studied were: nature of the rubber stopper; rubber stopper thickness, type of metal needle bevel used to pierce the stopper, and puncture technique. For each one of 16 different situations, 40 medication vial rubber stoppers were punctured, and the contents filtered. The filters were then examined under optical microscopy and particles present counted and measured.Results:The incidence of particle formation ranged from 0 % to 75 %, depending on the situation. Particle length was on average of 0.98±0.39 mm. The situation that gave the most particles (75 %; 30/40) was obtained when using a short bevelled needle, a 4 mm thick chlorobutyl vial stopper and with a puncture angle of 90°. Whilst a puncture technique reduced particle formation by more than 50 % for the most at risk situation, but without eliminating particle formation (residual formation of 22.5 %; 11/40), the use of a blunt bevelled needle totally eliminated the incidence of visible particle creation. The thickness of the rubber and the nature of the elastomer seemed to be linked to coring incidence, but in lesser proportions.Conclusion:Puncturing the stoppers using a technique with a 45° puncture angle reduced particle formation, but only the use of a blunt metal needle totally eliminated it.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.