The doses from the cobalt-60 teletherapy machines were measured using the FBX and secondary-standard dosemeters of Farmer-Baldwin type. The FBX dosemeter contained 0.20 mM ferrous ammonium sulphate, 5.0 mM benzoic acid and 0.20 mM xylenol orange in 0.05 N sulphuric acid. The values were compared with the values from a Fricke dosemeter and a graphite chamber used as primary standards. The values of the FBX, Fricke and graphite chambers agreed. There were, however wide differences among the different secondary-standard dosemeters themselves and with the FBX dosemeter. The FBX dosemeter was used for the measurement of central axis depth dose distributions for 5, 8, 10, 20 and 30 MeV electron and 42 MV x-ray beams.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.