The accuracy of the Ohmeda Biox 3700 pulse oximeter was evaluated in 10 paediatric patients, deliberately surface cooled to 25 degrees C in preparation for cardiac surgery, by comparing the arterial oxygen saturation results obtained from an Ohmeda Biox 3700 and a Radiometer OSM-2 Hemoximeter. Though there was a good correlation between the two series of results, the arterial oxygen saturation was over-estimated by the pulse oximeter compared to the OSM-2 Hemoximeter in the temperature range 36 degrees to 30 degrees C and under-estimated below 30 degrees C. These differences were greatest when the initial saturations were low.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.