AimsThe aim of the study was to present a single-centre experience of pacemaker and implantable cardioverter defibrillator (ICD) lead extraction using different methods, mainly laser-assisted extraction.Methods and resultsData from 1032 leads and 647 procedures were gathered. A step-by-step approach using different techniques while performing an ongoing risk–benefit analysis was used. The most common indications were local infection, systemic infection, non-functional lead, elective lead replacement, and J-wire fracture. Mean implantation time for all leads was 69 months and for laser-extracted leads 91 months. Laser technique was used to extract 60% of the leads, 29% were manually extracted, 6% extracted with mechanical tools, 4% were surgically removed, and 0.6% extracted by a femoral approach. Failure rate was 0.7%, and major complication rate was 0.9%. No extraction-related mortality occurred. Median time for laser extraction was 2 min. Long implantation time was not a risk factor for failure or for complication.ConclusionPacing and ICD leads can safely, successfully, and effectively be extracted. Leads can often be extracted by a superior transvenous approach; however, open-chest and femoral extractions are still required. Laser-assisted lead extraction proved to be a useful technique to extract leads that could not be removed by manual traction. The results indicate that the paradigm of abandoning redundant leads, instead of removing them, may have to be reconsidered.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.