SummaryWe assessed whether¯exible ®breoptic-guided orotracheal intubation could be rapidly and successfully achieved during a simulated rapid sequence induction in 30 anaesthetised and paralysed patients. Rapid sequence induction was simulated by applying practised cricoid pressure. Using a¯exible ®breoptic laryngoscope with camera and closed circuit television, an anaesthetist experienced with the technique performed orotracheal endoscopy and intubation with a cuffed 7.0-mm Portex tracheal tube through a VBM Bronchoscope Airway. Fibreoptic intubation was successful at the ®rst attempt in 28 patients (93%); two patients required two attempts. Mean (SD) time from removal of the facemask from the patient's face to the appearance of carbon dioxide in the expired breath after intubation was 111 (46) s (median 100 s; range 54±195 s). There were one or more dif®culties in 13 patients (43%). These dif®culties were largely avoidable and included problems with ®breoptic equipment, the Bronchoscope Airway, copious secretions, cricoid pressure or railroading of the tracheal tube. Flexible ®breoptic-guided orotracheal intubation may have a place in the management of failed intubation during a rapid sequence induction.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.