The purpose of this study was to develop an automatic method for validating the computed tomography gantry tilt. A head polymethyl methacrylate phantom with a diameter of 16 cm was used. Gantry tilt angles were measured both manually and automatically. Manual measurements were performed by measuring the length of the anteroposterior and lateral diameters from acquired images using electronic calipers. Automatic measurements consisted of a number of steps: phantom segmentation, determination of the center of the phantom, measurement of the anteroposterior and lateral diameters, and computation of the gantry tilt angle. The method was implemented on the gantry angles from 0° to 15°. The proposed method of measuring gantry angles produced accurate gantry tilt angles. The differences with the angles displayed on the gantry were less than 1°. The results of the automatic method were the same as those of the manual method (R2 > 0.98).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.