Micropumps are the most important components of lab-on-a-chip devices, which are becoming popular recently due to their enormous advantages. Among the different designs of micropumps, the nozzle-diffuser valveless design is the most preferred one due to simplicity in manufacturing. For the simulation of these pumps, the local fluid-structure interaction modeling is important to get the performance accuracy. To capture fluid-structure interaction accurately, an appropriate selection of the fluid as well as the structural modeling approaches are very essential. It is well known that the polydimethylsiloxane materials behave as a viscoelastic material. However, most of the earlier studies on polydimethylsiloxane micropump modeling have considered polydimethylsiloxane as a linear elastic material. In this article, a nozzle-diffuserbased valveless micropump is modeled using a fluid-structure interaction approach along with a linear viscoelastic model for polydimethylsiloxane. The mechanical properties of the polydimethylsiloxane material are experimentally obtained using dynamic mechanical analysis for the range of micropump operating frequencies. These viscoelastic properties of polydimethylsiloxane, with varying proportionate of curing agent, are introduced in the fluid-structure interaction model of micropump using the Kelvin-Voigt model. The results are presented in the form of micropump performance parameters such as diaphragm deflection, von-Mises stresses, swept volume, and fluid flow rate.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.