The various factors affecting energy dispersive X-ray fluorescence determination of iron in solid powder samples were evaluated, to improve the quality of these measurements. Influence of samples/standards preparation and instrumental parameters such as applied voltage to the X-ray tube, tube output current, and duration of acquisition were assessed, and the optimized instrumental parameters were 12 kV, 200 μA, and 200 s, respectively. Linear dynamic range of the spectrometer was found to be 0.0052 to 5.042 g kg −1 . Detection limits of iron in various organic/inorganic samples were found to be 2.5 to 21.9 mg kg −1 . The values obtained during the present method were evaluated with neutron activation analysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.