This study evaluated the extended use of a haematology analyser (Abbott Cell-Dyn CD4000) for the immunofluorescent enumeration of foeto-maternal haemorrhage (FMH) with fluorescein isothiocyanate-labelled monoclonal anti-RhD. Method performance was assessed with artificial FMH standards, and a series of 44 clinical samples. Within run precision was <15% (coefficient of variation, CV) for FMH volumes of 3 ml and above, 18.8% at an FMH volume of 2 ml and 31.7% at an FMH volume of 1 ml. Linearity analysis showed excellent agreement (observed FMH% = 0.98x expected FMH% + 0.02), and a close relationship (R(2) = 0.99) between observed and expected FMH percentages. The lower limit of quantification of the CD4000 (SRP-Ret) method with a maximum CV of 15% was 1.6 ml, and the limit of detection was <1 ml. Parallel Kleihauer-Betke test (KBT) assessments of FMH standards showed an overall trend for higher KBT values (observed = 1.25x expected - 0.38). At an FMH level of 4 ml, KBT observer estimates ranged from 0.57 to 11.94 ml with a mean inter-observer CV of 63%. For 44 clinical samples, there was decision point agreement between KBT and SRP-Ret results for 42 samples with an FMH of <2 ml. Analysis in the low FMH range (<1 ml) showed that small volume foetal leaks could be detected with the SRP-Ret method in most of 23 samples with negative KBT results. CD4000 SRP-Ret method performance for FMH determination was similar to that reported for flow cytometry.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.