Implantation of a subcutaneous array to improve the defibrillation threshold of an existing transvenous defibrillation lead system without the need for lead extraction is discussed.
(J Cardiovasc Electrophysiol, Vol. 14, pp. 314‐315, March 2003)
Occasional patients have excessive defibrillation energy requirements despite appropriate transvenous defibrillation lead position and modification of defibrillation waveform and configuration. Preliminary data suggest that use of subcutaneous defibrillation electrode arrays with nonthoracotomy systems is associated with a substantial reduction in defibrillation threshold. The current operative approach to subcutaneous lead array implantation involves the use of a separate left chest incision. We present two cases in which implantation of a subcutaneous lead array in combination with a transvenous defibrillation electrode was performed via a single infraclavicular incision and associated with a reduction in defibrillation threshold. Such an approach simplifies implantation and avoids the potential morbidity of the additional incision required of a left lateral chest approach.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.