We have studied the progression of healing in 103 unstable fractures of the tibia. In 76 patients we removed the external fixator once the stiffness had reached 15 Nm/° in the sagittal plane. Deformity at the site of the fracture subsequently occurred in four patients. In a further 27, we measured stiffness in several planes and removed the fixator only when the stiffness reached 15 Nm/° in each. We found that stiffness in two orthogonal planes may differ widely (maximum difference 9.0 Nm/°, mean 4.1 Nm/°). There were no failures in the second group. We advocate that fracture stiffness be measured in two orthogonal planes when assessing tibial healing and suggest that values above 15 Nm/° in two planes give an indication that it is safe to remove the fixator.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.