In this work, we developed a HPLC method for the multidetermination of avermectins (AVM) (abamectin-ABA 1b and ABA 1a, eprinomectin-EPR, and ivermectin-IVM) and milbemycins (moxidectin-MOX) in milk samples using polypyrrole (PPy) as adsorbent material in pipette-tip solid-phase extraction (PT-PPy-SPE). PPy was characterized by scanning electron microscopy, thermogravimetric analysis, Fourier transform infrared spectroscopy, and X-ray diffraction and the data agreed with the literature. The sample preparation included the clean-up of the milk by protein precipitation (PP) with acetonitrile and extraction of the analytes by PT-PPy-SPE. The chromatographic method was developed in reverse phase and isocratic mode with flow rate at 1.2 mL min and ultraviolet detection at 250 nm. The mobile phase composition was acetonitrile:methanol:water (55:25:20, v/v/v). The studied parameters and the optimized conditions for the sample preparation were washing solvent (300 μL water), volume and type of eluent (500 μL methanol), volume and pH of sample (1 mL and pH 10), amount of adsorbent material (50 mg PPy), and without addition of salt (NaCl). The method was linear over the concentration range from 20 to 3000 ng mL with coefficients of correlation (r) ≥ 0.99 for all analytes and recoveries around 100%. The method developed and validated was used for the analyses of real milk samples from cow treated with Ivomec (IVM 3.5%), in which were found 21.51 ± 2.94 ng mL of IVM. Finally, the results proved that PT-PPy-SPE coupled to HPLC-UV was economical, simple, and easy-to-perform technique. Graphical abstract Pipette-tip solid phase extraction using polypirrole as adsorbent material for determination of avermectins and milbemycins in milk.
This work aims to develop analytical methods using high-performance liquid chromatography with a diode array detector (HPLC-DAD) for analysis and quantification of avermectins (AVMs) and milbemycins (MBMs) in bulk samples. First, the methods were optimized and some parameters such as temperature, flow rate, injection volume and mobile phase with different proportions of solvents were evaluated. The best chromatographic conditions were obtained using the Phenomenex ® C18 (150 × 4.60 mm, 5 μm) column at a temperature of 20 °C, flow rate of 1.2 mL min -1 , injection volume of 20 μL, and detection at 250 nm. Acetonitrile: ultrapure water (87: 13, v/v) was used as mobile phase for moxidectin and eprinomectin, and acetonitrile: methanol: ultrapure water (53: 35: 12, v/v/v) for abamectin and ivermectin. Under these conditions satisfactory results were obtained, with appropriate limits of detection and quantification, acceptable linearity, precision, accuracy, and robustness. These methods satisfy the need for analytical methods for the multi-determination of MBMs and the B1a and B1b forms of AVMs by HPLC-DAD, which can be considered simple, effective, innovative and should aid in the development of the field.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.