A quadrupole/orbital trapping mass
spectrometer or Q-Exactive (QE)
interfaced with a gas chromatograph (GC) was optimized for measuring
polychlorinated dibenzo-p-dioxins, dibenzofurans
(PCDD/Fs), and polychlorinated biphenyls (PCBs) in foods. Figures
of merit include (1) an instrument detection limit (IDL) for 2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD) of 9 femtograms (fg), (2) quantitative mass
resolution from PCDD interferences (e.g., PCBs, methoxy-PCBs DDTs,
polychlorodibenzylphenyl ethers, polychloroxanthenes, methyl-polychlorodibenzofuran,
and polychlorodibenzothiophenes), and (3) mass accuracy <1 ppm
at the IDL. The QE measured the concentrations of PCDD/Fs and PCBs
in whole cow’s milk with no known source of contamination (e.g.,
TCDD 33 fg/g fat). A National Institute of Standards and Technology
(NIST) unfortified human milk standard reference material (SRM) 1953
was measured determining 27 PCDD/F and PCB congeners with an average
difference of 7.6% from the certified results. The QE-GC is a benchtop
instrument, easy to service, easy to operate, and requires no lock
masses, mass preselection, or chemical ionization conditions. The
QE-GC demonstrated that it can be an alternative to the double focusing
magnetic sector instruments (sector) for the high-resolution measurement
of PCDD/Fs and PCBs in dairy products.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.