Background: A point-of-care device that can provide immediate and reliable hemoglobin (Hb) concentrations and packed cell volumes (PCVs) would be useful in veterinary medicine.
Objectives:We aimed to compare the use of a human device (Mission Plus; MP) with a gold standard (GS) method for measuring Hb concentrations and PCVs in cattle blood.Methods: Blood samples from clinically healthy cattle (n = 122) were collected with or without an anticoagulant (K 2 EDTA). The GS and MP methods were compared with correlation coefficients. Passing-Bablok regression analyses were also performed, and the acceptability judgment was completed using Bland-Altman plots.
Results:The CV max for Hb values obtained using the GS method, the MP device without K 2 EDTA, and the MP device with K 2 EDTA were approximately 2.70%, 1.70%, and 2.0%, respectively, whereas the CV max for PCVs was 0.90%, 1.83%, and 2.05%, respectively. A positive correlation (97.5% confidence interval) was observed between the Hb concentrations and PCV values detected using the MP and GS techniques in blood with and without K 2 EDTA. Bland-Altman plots showed agreement between the MP and GS methods. For Hb using blood collected with or without the addition of K 2 EDTA, the mean differences were −0.87 g/dL (95% CI: 1.35; −3.96) and 0.08 g/ dL (95% CI: 2.16, −1.99), respectively. For PCVs using blood collected with or without the addition of K 2 EDTA, the mean differences were −3.75% (95% CI: 0.61. −8.12) and −0.88% (95% CI: 2.86, −4.62), respectively.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.