Micro-finite element (μFE) analysis has recently been introduced for the detailed quantification of the mechanical interaction between bone and implant. The technique has been validated at an apparent level. The aim of this study was to address the accuracy of μFE analysis at the trabecular level. Experimental displacement fields were obtained by deformable image registration, also known as strain mapping (SM), of dynamic hip screws implanted in three human femoral heads. In addition, displacement fields were calculated using μFE analysis. On a voxel-by-voxel basis, the coefficients of determination (R(2)) between experimental and μFE-calculated displacements ranged from 0.67 to 0.92. Linear regression of the mean displacements over nine volumes of interest yielded R(2) between 0.81 and 0.84. The lowest R(2) values were found in regions of very small displacements. In conclusion, we found that peri-implant bone displacements calculated with μFE analysis correlated well with displacements obtained from experimental SM.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.