Introduction: Reliability of biological assays results depends closely on the quality of the 3 biological stages: pre-analytical, analytical, and post-analytical. In each stage, a quality program including internal quality control (IQC) and external quality assurance (EQA) has to be assessed to ensure test result accuracy and precision. The coagulation laboratory plays an important role in the diagnosis and treatment of people with blood disorders, so though several lack of facilities in Malagasy labs, QC has to be implemented. Thus, this study aimed to analyze IQC and EQA in hemostasis tests in the Hematology laboratory of the CHU-HJRA in Antananarivo Madagascar in order to improve its analytical phases quality. Methods: This retrospective descriptive and analytical study was conducted from January 1st, to October 31st, 2021 in the Hematology laboratory of CHU-JRA within the routine coagulation tests (Prothrombin time PT and Activated partial prothrombin time aPTT). Regarding the EQA, percentage of deviation from the peer groups results were calculated. Then in order to analyze IQCs results, coefficient of variation and biases of IQCs results, corresponding respectively to precision and accuracy, were calculated and Levey-Jennings graphs were analyzed. Results: Regarding global results of EQA, 50% were within consensus compared to peer-group. And regarding IQC, for precision, one coefficient of variation for both PT and aPTT exceeded 5%, and for accuracy, no bias exceeds 10% deviation from the expected result. Levey-Jenningss plot showed more than 90% of the results to be within M±2 SD range. A violation 10x Westgard rules, corresponding to a systematic error was observed. Conclusion: Each out-of-control results has to be analyzed as part of strengthening quality program guaranteeing coagulation tests reliability.
Objective: to compare the two immunological methods of HbA1c determination available at the two university hospital centers in the capital of Madagascar. Methods:Prospective comparative study between two HbA1C assay techniquesat the University Hospital Center Joseph RavoahangyAndrianavalona and University Hospital Center Joseph RasetaBefelatanana both localized at Antananarivo Madagascar was conducted. The two techniques used are immunoturbidimetry and immunofluorescence. Fifty samples were run simultaneously on both techniques to assess the dispersion and correlation of HbA1C values. A history of haemoglobin abnormalities is previously researched in the medical records for the samples. Results:The correlation coefficient for the 2 techniques is 0.97. The Bland –Altman plot shows a funnel-shaped appearance. The Interclass Correlation Coefficient is 0.851 CI [-0.036 - 0.962]. No hemoglobin abnormalities were documented in any of the 50 samples Conclusion: It is essential to keep the same technique for monitoring HbA1c. Our study shows poor reproducibility for the two techniques compared. A strong relationship between the technique used and the HbA1c value was found for the extreme values. It is important to specify the technique used to support prescribers.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.