We design and build a photometric ellipsometer that can be adjusted or modified to match specific needs for different experiments. To validate our setup, we test the system with glass substrates at multiple incident angles from 30°to 70°. The experimental data can then be fitted to the standard theoretical model for ellipsometry with a single interface, allowing the amplitude ratio (tan W) and the phase difference (D) to be evaluated as fitting parameters. As a result, we obtain the refractive indices of glass as 1.44-1.55 depending on the backing material. By the same means, it is also feasible to derive thicknesses of silicon oxide films. The resulting thicknesses are in good agreement with those determined by a commercial ellipsometer with the minimum deviation of 0.2 %.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.