Aims:To evaluate the accuracy of measurement of hepatic venous pressure gradient by catheter wedge as compared to balloon wedge (the gold standard).Materials and Methods:Forty-five patients having a clinical diagnosis of intrahepatic portal hypertension were subjected to the two different types of pressure measurements (catheter wedge and balloon wedge) during transjugular liver biopsy under fluoroscopic guidance.Statistical Analysis:Spearman's rank correlation coefficient, Bland–Altman plot for agreement, and single measure intraclass correlation were used for analysis of data.Results:There was a close correlation between the results obtained by both the techniques, with highly significant concordance (P < 0.0001). Hepatic venous pressure gradients as measured by the catheter wedge technique were either equal to or less than those obtained by the balloon wedge technique.Conclusions:The difference in hepatic venous pressure gradients measured by the two techniques is insignificant.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.