Optical low-coherence reflectometry (OLCR) was used as a noncontact method to measure the central corneal thickness of three patients intraoperatively during photorefractive keratectomy. Continuous on-line measurements were performed on the intact cornea immediately before the beginning of surgery, after the removal of the corneal epithelium, during laser tissue photoablation, and for 3 min after the ablation process. Corneal thinning due to evaporation was studied on a separate patient with the OLCR instrument, and it was found to be -0.14 microm/s during the first 5 min after epithelium removal. This baseline corneal thinning rate was used as a fit parameter to calculate actual from measured ablation depths. The measurements showed a maximum difference of +/-10 microm between planned ablations (34-92 microm) and measured ablation depths.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.