Advances in surgical technique and implant technology have improved the ten-year survival after primary total hip arthroplasty (THA). Despite this, the number of revision procedures has been increasing in recent years, a trend which is predicted to continue into the future. Revision THA is a technically demanding procedure often complicated by a loss of host bone stock which may be compounded by the need to remove primary implants. Both cemented and uncemented implant designs are commonly used in the United Kingdom for primary and revision THA and much controversy still exists as to the ideal method of stem fixation. In this article we discuss revision of the femur using cemented components during revision THA. We focus on three clinical scenarios including femoral cement-in-cement revision where the primary femoral cement-bone interface remains well fixed, femoral cement-in-cement revision for peri-prosthetic femoral fractures, and femoral impaction grafting. We discuss the clinical indications, surgical techniques and clinical outcomes for each of these procedures.
We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up. The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips. All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure. Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p < 0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p < 0.001). The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis. Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.
This is the largest series of revision total hip arthroplasties with femoral impaction grafting, and the results support the continued use of this technique. Cite this article: Bone Joint J 2016;98-B:1611-19.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.