Background
Gastric electric stimulation (GES) at a high-frequency, low-energy setting is an option for treating refractory gastroparesis. The currently available commercial stimulator, the Enterra neurostimulator (Medtronic Inc, Minneapolis, MN), however, requires surgical implantation and is powered by a nonrechargeable battery.
Objective
To develop and test a miniature wireless GES device for endoscopic implantation in an experimental model.
Design
In-vivo gastric signals were recorded and measured in a nonsurvival swine model (n = 2; 110-lb animals).
Intervention
An endoscopically placed, wireless GES device was inserted into the stomach through an overtube; the two GES electrodes were endoscopically attached to the gastric mucosa and secured with endoclips to permit stimulation.
Main Outcome Measurements
Stable electrogastrogram measures were observed during GES stimulation.
Results
Electrogastrogram recordings demonstrated that gastric slow waves became more regular and of constant amplitudes when stomach tissues were stimulated, in comparison with no stimulation. The frequency-to-amplitude ratio also changed significantly with stimulation.
Limitation
Nonsurvival pig studies.
Conclusion
Gastric electric stimulation is feasible by our endoscopically implanted, wireless GES device.
Background
Gastric stimulation via high-frequency, low-energy pulses can provide an effective treatment for gastric dysmotility; however, the current commercially available device requires surgical implantation for long-term stimulation and is powered by a nonrechargeable battery.
Objective
To test and describe endoscopic implantation techniques and testing of stimulation of a novel, wireless, batteryless, gastric electrical stimulation (GES) device.
Design
Endoscopic gastric implantation techniques were implemented, and in vivo gastric signals were recorded and measured in a non-survival swine model (n = 2; 50-kg animals).
Intervention
Five novel endoscopic gastric implantation techniques and stimulation of a novel, wireless, batteryless, GES device were tested on a non-survival swine model.
Main Outcome Measurements
Feasibility of 5 new endoscopic gastric implantation techniques of the novel, miniature, batteryless, wireless GES device while recording and measurement of in vivo gastric signals.
Results
All 5 of the novel endoscopic techniques permitted insertion and securing of the miniaturized gastrostimulator. By the help of these methods and miniaturization of the gastrostimulator, successful GES could be provided without any surgery. The metallic clip attachment was restricted to the mucosal surface, whereas the prototype tacks, prototype spring coils, percutaneous endoscopic gastrostomy wires/T-tag fasteners, and submucosal pocket endoscopic implantation methods attach the stimulator near transmurally or transmurally to the stomach. They allow more secure device attachment with optimal stimulation depth.
Limitations
Non-survival pig studies.
Conclusion
These 5 techniques have the potential to augment the utility of GES as a treatment alternative, to provide an important prototype for other dysmotility treatment paradigms, and to yield insights for new technological interfaces between non-invasiveness and surgery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.