Introduction. Noninvasive blood pressure (NIBP) monitoring methods are widely used in critically ill patients despite poor evidence of their accuracy. The erroneous interpretations of blood pressure (BP) may lead to clinical errors. Objectives. To test the accuracy and reliability of aneroid (ABP) and oscillometric (OBP) devices compared to the invasive BP (IBP) monitoring in an ICU population. Materials and Methods. Fifty adult patients (200 comparisons) were included in a randomized crossover trial. BP was recorded simultaneously by IBP and either by ABP or by OBP, taking IBP as gold standard. Results. Compared with ABP, IBP systolic values were significantly higher (mean difference ± standard deviation 9.74 ± 13.8; P < 0.0001). Both diastolic (−5.13 ± 7.1; P < 0.0001) and mean (−2.14 ± 7.1; P=0.0033) IBP were instead lower. Compared with OBP, systolic (10.80 ± 14.9; P < 0.0001) and mean (5.36 ± 7.1; P < 0.0001) IBP were higher, while diastolic IBP (−3.62 ± 6.0; P < 0.0001) was lower. Bland-Altman plots showed wide limits of agreement in both NIBP-IBP comparisons. Conclusions. BP measurements with different devices produced significantly different results. Since in critically ill patients the importance of BP readings is often crucial, noninvasive techniques cannot be regarded as reliable alternatives to direct measurements.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.