The correction of repositioning along the AP and SI direction from conventional bone matching in CT image-guided proton therapy was found to be effective to maintain the dose constraint of the rectum and the dose coverage of the prostate. This work indicated that prostate cancer treatment by prostate matching using CT image guidance may be effective to reduce the rectal complications and achieve better tumor control of the prostate. However, an adaptive approach is desirable to maintain better dose coverage of the SVs.
The positional deviations for the prostate on the posterior side and the SVs were smaller by PRB matching than the other strategies and effectively reduced the rectal dose. 3D dose calculations indicate that PRB matching with CT image guidance may do a better job relative to other positioning methods to effectively reduce the rectal complications. The WEL variation was quite large, and the appropriate margin (approx. 10 mm) must be adapted to the proton range in an initial planning to maintain the coverage of target volumes throughout entire treatment.
The efficacy of proton beam therapy (PBT) for hepatocellular carcinoma (HCC) has been reported, but insertion of fiducial markers in the liver is usually required. We evaluated the efficacy and toxicity of respiratory-gated PBT without fiducial markers for HCC located within 2 cm of the gastrointestinal tract. From March 2011 to December 2015 at our institution, 40 patients were evaluated (median age, 72 years; range, 38–87 years). All patients underwent PBT at a dose of 60 to 80 cobalt gray equivalents (CGE) in 20 to 38 fractions. The median follow-up period was 19.9 months (range, 1.2–72.3 months). The median tumor size was 36.5 mm (range, 11–124 mm). Kaplan–Meier estimates of the 2-year overall survival, progression-free survival, and local tumor control rates were 76%, 60%, and 94%, respectively. One patient (2.5%) developed a grade 3 gastric ulcer and one (2.5%) developed grade 3 ascites retention; none of the remaining patients developed grade >3 toxicities (National Cancer Institute Common Terminology Criteria for Adverse Events ver. 4.0.). This study indicates that PBT without fiducial markers achieves good local control without severe treatment-related toxicity of the gastrointestinal tract for HCC located within 2 cm of the gastrointestinal tract.
We evaluated the effectiveness and toxicity of proton beam therapy (PBT) for hepatocellular carcinomas (HCC) >5 cm without fiducial markers using four-dimensional CT (4D-CT) planning. The subjects were 29 patients treated at our hospital between March 2011 and March 2015. The median total dose was 76 Cobalt Gray Equivalents (CGE) in 20 fractions (range; 66–80.5 CGE in 10–32 fractions). Therapy was delivered with end-expiratory phase gating. An internal target volume (ITV) margin was added through the analysis of respiratory movement with 4D-CT. Patient age ranged from 38 to 87 years (median, 71 years). Twenty-four patients were Child–Pugh class A and five patients were class B. Tumor size ranged from 5.0 to 13.9 cm (median, 6.9 cm). The follow-up period ranged from 2 to 72 months (median; 27 months). All patients completed PBT according to the treatment protocol without grade 4 (CTCAE v4.03 (draft v5.0)) or higher adverse effects. The two-year local tumor control (LTC), progression-free survival (PFS), and overall survival (OS) rates were 95%, 22%, and 61%, respectively. The LTC was not inferior to that of previous reports using fiducial markers. Respiratory-gated PBT with 4D-CT planning without fiducial markers is a less invasive and equally effective treatment for large HCCs as PBT with fiducial markers.
The distances and directions of respiratory motion differed for each liver segment, and a history of abdominal surgery reduced the respiratory motion of the liver. It is necessary to selectively use the internal margin setting.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.